Abstract:Objective: To evaluate the image quality, diagnostic accuracy and radiation doses for coronary artery imaging with second-generation 128-slice dual-source CT (DSCT) using the highpitch mode. Methods: A total of 123 cases patients with suspected or known coronary artery disease were randomly divided into two groups. Group A (n=58) underwent the adaptive prospective ECG-triggered sequence scan (CorAdpSeq) and group B (n=65) underwent the prospective high-pitch mode (pitch 3.4, Flash mode). Two experienced radiologists assessed the image quality and significant coronary stenosis (≥50%) with a double blinded fashion in a five score scale. Interobserver agreement for image quality was calculated with Cohen kappa statistics. The radiation doses values were measured. The diagnostic performance presented by DSCT in detecting stenosis ≥50% were compared with the gold standard DSA. Results: There were 1 968 segments of coronary arteries in 123 patients included in this analysis. Diagnostic image quality (score≥3) was obtained in 1920 segments(97.6%) of the coronary artery segments, which was obtained in 98.1%(910/928)of segments in group A and 97.1%(1 010/1 040)in group B respectively with no significant difference (χ2=0.030,P=0.867). The mean image quality score for the whole coronary segments was (4.7±0.8) in 123 patients.There was no statistical difference between mean image quality of group A (4.7±0.7) and group B (4.6±0.9) (t=0.603,P=0.548). The kappa value between two radiologists was 0.799(P<0.01)on image quality. The effective radiation dose in group B (1.08±0.11) mSv was significantly (t=11.642,P=0.000) lower than that in group A (2.26±0.49) mSv. Sensitivity and specificity were 97.6% and 97.5% pervessel in group A; 98.1%,96.5% per-vessel in group B. Conclusion: Compared with DSA both the high-pitch and CorAdpSeq mode for CTCA provide high image quality and accuracy of significant coronary stenosis, but the high-pitch mode can significantly reduce the radiation dose in the lower heart rate and regular heart rhythm.
收稿日期: 2013-04-20
通讯作者:
潘昌杰(通讯作者),副主任医师,Email:pcj424815@sina.com
作者简介: 吴春华(1973—),男,主管技师;
引用本文:
吴春华, 薛跃君, 许轶群, 荣伟良, 潘昌杰, 钱农. 双源CT大螺距扫描技术在冠状动脉狭窄诊断中的应用[J]. 江苏大学学报:医学版, 2013, 23(4): 345-349.
WU Chun-hua, XUE Yue-jun, XU Yi-qun, RONG Wei-liang, PAN Chang-jie, QIAN Nong. Dual-source CT coronary angiography using the high-pitch mode in diagnosing the coronary artery stenosis. Journal of Jiangsu University(Medicine Edition), 2013, 23(4): 345-349.
[1]王小玲, 赵青, 佟海滨, 等. 第二代双源CT前门控大螺距冠状动脉成像中ECG“不稳定波形”对图像质量的影响[J]. 中国医学影像技术, 2011, 27(11):2335-2339.[2]张艳艳, 冯博, 李佩玲, 等. 双源CT冠状动脉成像评价冠状动脉狭窄[J]. 中国医学影像技术, 2011, 27(8):1587-1590.[3]Leber AW, Johnson T, Becker A, et a1. Diagnostic accuracy of dual-source multi-slice CT-coronary angiography in patients with an intermediate pretest likelihood for coronary artery disease[J]. Eur Heart J, 2007, 28(19):2354-2360.[4]Scheffel H,Alkadhi H, Plass A, et a1. Accuracy of dual-source CT coronary angiography: first experience in a high pre-test probability population without heart rate control[J]. Eur Radiol, 2006, 16(12):2739-2747.[5]Brenner DJ, Hall EJ. Computed tomography: an increasing source of radiation exposure[J]. N Engl J Med, 2007, 357(22):2277-2284.[6]Herzog BA, Wyss CA, Husmann L, et al. First head-to-head comparison of effective radiation dose from low-dose 64-slice CT with prospective ECG-triggering versus invasive coronary angiography[J]. Heart, 2009, 95(20):1656-1661.[7]Scheffel H, Alkadhi H, Leschka S, et al. Lowdose CT coronary angiography in the step-and-shoot mode: diagnostic performance[J]. Heart, 2008, 94(9):1132-1137.[8]Lell M, Marwan M, Schepis T, et al. Prospectively ECG-triggered high-pitch spiral acquisition for coronary CT angiography using dual source CT: technique and initial experience[J]. Eur Radiol, 2009, 19(11):2576-2583.[9]Austen WG, Edwards JE, Frye RL, et a1. A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc Committee for Grading of Coronary Artery Disease, Council on Cardiovascular Surgery, American Heart Association[J]. Circulation, 1975, 51(Suppl 4):5-40.[10]Nakayama Y, Awai K, Funama Y, et a1. Low tube voltage reduces contrast material and radiation doses on 16-MDCT aortography[J]. AJR, 2006, 187(5):W490-W497.[11]Valentin J, International Commission on Radiation Protection. Managing patient dose in multidetector computed tomography(MDCT). ICRP Publication 102[J]. Ann ICRP, 2007, 37(1):1-79.[12]Deetjen A, Mollmann S, Conradi G, et a1. Use of automatic exposure control in multislice computed tomography of the coronaries: comparison of 16-slice and 64-slice scanner data with conventional coronary angiography[J]. Heart, 2007, 93(9):1040-1043.