Loss of capture caused by ventricular electrode’s micro dislocation: avoid electrode relocation by resetting pacemaker’s parameters |
|
|
Abstract For the pacemaker whose ventricular channel has automatic threshold management function and beat-by-beat monitoring function, if ventricular electrode’s micro dislocation leads to the increase of the pacing threshold, the automatic threshold management function can provide a high-output backup pulse in time when ventricular loss of capture occurs for initial pulse. This function can avoid the risk of ventricular dropped beats and infection of pacemaker pocket due to ventricular electrode’s relocation.
|
|
|
|
|